Adjusting an M&W level

Adjusting an M&W level

Home Forums Workshop Tools and Tooling Adjusting an M&W level

Viewing 17 posts - 1 through 17 (of 17 total)
  • Author
    Posts
  • #453024
    Robin King
    Participant
      @robinking15611

      I need to adjust a Moore and Wright ELS level which is noticeably out of calibration, and while familiar with the process for checking and adjusting I can't find any specific detailed instructions for the adjustment of this particular level. Like the ELM it appears to be provided with pairs of screws at both ends of the main vial tube, running vertically down into the base casting but it's not clear as to whether these are simply fixings or for adjustment. Before I dive in and do the wrong thing I thought for once I'd ask first

      Has anyone any information on these levels, please?

      #19633
      Robin King
      Participant
        @robinking15611
        #453029
        Martin Kyte
        Participant
          @martinkyte99762

          Is this any help.

          https://www.bowersgroup.co.uk/media/wysiwyg/MW550-03_Cropped.pdf

          I would have thought that the level adjuster(s) should be at 90 deg to the hold down screws and should act by inserting or withdrawing a tapered seating. You don't need a perfectly level surface to test it so long as it reads somewhere withing the scale. Just mark the initial position and rotate by 180 degrees for a second reading. They should be identical when perfectly adjusted.

          regards Martin

          #453033
          Howard Lewis
          Participant
            @howardlewis46836

            The technique that I used was (Admittedly using a surface table in the standards Room ) was to use one of the adjusting screws so that the reading was the same when the level had been rotated through 180 degrees.

            i e. if the bubble just touches one of the graduation lines when facing North, rotate to face South and adjust by half the difference.

            Return to face North, again and if there is any difference, move the same adjusting screw to halve the difference.

            Keep repeating this sequence of operations until the reads the same which ever way the level is facing. Lock the adjuster, and repeat checks, to ensure that locking has had no effect. If it has affected the setting, return to start of sequence and repeat the process until there is no change in reading.

            Howard

            #453036
            Robin King
            Participant
              @robinking15611

              Thank you both.

              Martin, unfortunately not, as that level is an entirely different form of construction to the British made ELS/ELM types.

              Howard, yes indeed, that's exactly the method that I use but my problem with the ELS is in identifying which of the screws do what on this type.

              #453039
              martin perman 1
              Participant
                @martinperman1

                I've just had a look at a M&W els level and I can only see a vertical screw at one end and a horizontal screw at the other, am I correct.

                Martin P

                #453040
                martin perman 1
                Participant
                  @martinperman1

                  Sorry

                  Edited By martin perman on 19/02/2020 16:27:03

                  #453041
                  Howard Lewis
                  Participant
                    @howardlewis46836

                    Since I don't know the M & W ELS level, so afraid that I cannot be much help.

                    FWIW I would scribe a mark on a vertical adjuster and then try adjusting that. If that doesn't work, at least you can return it to the original, scribed line, setting, before doing the same on one of the other adjusters.

                    Any chanced that you could post a picture of the adjusters at one end of the level?

                    Hopefully, someone will be able to give some better advice.

                    Howard

                    #453045
                    Howard Lewis
                    Participant
                      @howardlewis46836

                      Having just looked at the pictures of ELS level on the Cromwell Tool site, it looks as if there are two grub screws at each end of the vial. The outer one may lock the end fitting, whilst the inner one may be a clamp for the vial. Hopefully these are just nipped enough to prevent any movement. Would not overtighten any of them for fear of cracking the vial.

                      The large horizontal one at one end looks to act as a pivot, and the large vertical screw at the other end is the one that I would attack.

                      If the vial is loose, it needs to be rotated on its axis so that the bubble goes to the highest point, before nipping up the clamp screw.. The vial may either be a curved tube, or it could possibly even be straight externally but ground to a curve internally.

                      If the vial is loose and needs to be rotated, you may need to slacken the outer grubscrew at the end with the vertical adjuster. removing the vertical adjuster screw would probably allow the end "plug" to be unscrewed to give access to the end of the vial. Once you are happy with the vial orientation, the "plug" can be refitted, locked with the grubscrew, and the vertical adjuster refitted.

                      Once you are happy with that, adjusting the vertical screw should just be a matter of iteration. Hopefully it will not be so slack that it moves easily to upset things again.

                      Hope that this is the right way to go about things.

                      Howard

                      fat fingers and useless proof reader!

                      Edited By Howard Lewis on 19/02/2020 16:49:49

                      #453048
                      Robin King
                      Participant
                        @robinking15611

                        Howard thanks. I'm fairly certain that the two screws at either end of the main vial are the ones needed for adjustment.

                        What appear to be larger slotted screw heads on top at one end and on the side at the other are repeated underneath and on the back, and as far as I can see are just end caps to the vertical vial and short cross vial, and not related to adjustment, hence my original query. I don't think I should fiddle with those.

                        R

                        #453050
                        martin perman 1
                        Participant
                          @martinperman1

                          It would appear to me that the two vertical screws top and bottom would be the ones to adjust, dependant on wether you raise or lower that end by releasing one and adjusting the other.

                          Martin P

                          Edited By martin perman on 19/02/2020 17:09:40

                          #453052
                          Graham Meek
                          Participant
                            @grahammeek88282

                            There is an exploded drawing of these levels in the M&W Spare Parts list, page 17 that I think will help.

                            Regards

                            Gray,

                            #453059
                            Robin King
                            Participant
                              @robinking15611

                              Graham, thanks – that's a great help; pity it doesn't give any clues as to how or if that's what you use to adjust the main vial. I'm reluctant to experiment!

                              Robin

                              #453138
                              David Davies 8
                              Participant
                                @daviddavies8

                                Robin

                                I was recently given an ELM level which is a 12" version of the ELS. It had a broken vial. I removed the old vial which was held in by a potting compound similar to Polyfilla and am about to rebuild it with a vial obtained at a very reasonable price from Cromwell tools (I have no connection). Each end of the vial holding tube has two vertical screws, one screwhead inside the end plug of the tube, the other beneath the end of the tube. both have access holes for an allen key.The lower of the two screws acts as a jack for calibration, the upper one is a securing screw. To calibrate, the securing screw is released and the jacking screw is adjusted as required. I haven't done this yet but my understanding is that the level is calibrated by obtaining the same deviation when the level is swapped end for end on a flat surface. Once done, possibly over many iterations, the securing screw is tightened and a final check is made.

                                HTH

                                Dave

                                #453144
                                Clive Hartland
                                Participant
                                  @clivehartland94829

                                  Yes, the term is, 'End for End. when adjusting it means that you need a stable base like a lathe bed, or a surface plate. determine the error, end for end and see result, bubble up, screw down and so forth. It is just a matter of adjusting the table your using as well. until bubble settles in both directions when end for ended.

                                  #453169
                                  Robin King
                                  Participant
                                    @robinking15611

                                    Dave, many thanks, that's just the info I was after and makes sense on looking at it again so I'll work on that basis.

                                    Clive, thanks. The method I was shown years ago was to use a surface plate but to rotate the level until it centres correctly, which it must do at 90 degrees to the slope (if any), place an angle plate alongside it held in place to use as a lateral reference, then turn the level 180 degrees and place alongside the angle plate to check if still centred. Adjust the level as necessary; no need to adjust the surface plate.

                                    #453170
                                    Robin King
                                    Participant
                                      @robinking15611

                                      Obviously re-check and repeat the process until the error is corrected.

                                    Viewing 17 posts - 1 through 17 (of 17 total)
                                    • Please log in to reply to this topic. Registering is free and easy using the links on the menu at the top of this page.

                                    Latest Replies

                                    Viewing 25 topics - 1 through 25 (of 25 total)
                                    Viewing 25 topics - 1 through 25 (of 25 total)

                                    View full reply list.